Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Beckman AC30A: (1 year) CAL VER /5500 DATE: 27-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.6 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 7 NUMBER OF LINES: 70 CONFIGURATION: Fluke 5500A STANDARD: Fluke 5500A/COIL ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # acdv.bmp # acd50.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F V 1.002 ASK+ K X 1.003 STD Fluke 5500A/COIL 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: NONE 1.005 DISP Temperature: 18 - 28 degC 1.005 DISP Humidity: <80% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD {AC VOLTAGE} 1.008 DISP Set the POWER/RANGE switch to AC VOLTS. 1.009 PIC acdv 1.010 HEAD AC VOLTAGE: {600V Range} 1.011 5500 600 500V 1% 5U 60H SI 2W 2.001 ASK+ R 2.002 HEAD {RESISTANCE} 2.003 DISP Connect the 5500A to the UUT as follows: 2.003 DISP [32] 5500A AUX HI to UUT V/OHM 2.003 DISP [32] 5500A AUX LO to UUT COM 2.003 DISP [32] 5500A NORMAL HI to UUT V/OHM 2.003 DISP [32] 5500A NORMAL LO to UUT COM 2.003 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 2.003 DISP the NORMAL leads to the SENSE leads AT the UUT. 2.004 RNG 2kZ 2.005 HEAD RESISTANCE: {2 kOhm Range} 2.006 5500 0.0Z S CW 2.007 EVAL Is an audible tone present {(Continuity check OK)}? 3.001 5500 2 1.000kZ 1% 0.002U CW 4.001 HEAD {AC CURRENT} 4.002 PIC acd50 4.003 HEAD AC CURRENT: {20A Range} 4.004 5500 20 10.00A 2% 0.07U 60H SI T5 5.001 HEAD AC CURRENT: {200A Range} 5.002 5500 200 100.0A 1.5% 0.7U 60H SI T5 6.001 HEAD AC CURRENT: {400A Range} 6.002 5500 400 300A 1.5% 5U 60H SI T5 7.001 END